Product Short Description

Product Name​
NI SCXI-1130 256-Channel Reed Matrix / Multiplexer Switch Module
Product Introduction​
The NI SCXI-1130 is a high-density SCXI switch module. It can operate as a multiplexer, matrix or independent switch bank for routing many signals to instruments or DAQ channels.

Description

Technical Specifications​
Part number:​ 778661-01
Maximum channels:​ 256
Multiplexer modes:​ 256×1 single-wire, 128×1 two-wire, 64×1 four-wire, 16 groups of 16×1
Matrix modes:​ 4×64 single-wire, 8×32 single-wire, 4×32 two-wire
Maximum switching voltage:​ 60 VDC
Maximum switching current:​ 0.4 A
Maximum switching power:​ 10 W per channel
Relay type:​ reed relay
Connectors:​ two 176-pin docking station plug I/O connectors
Dimensions:​ 1.2 x 6.9 x 8.0 in.
Weight:​ 1.020 kg
Operating temperature:​ 0 to 55 °C
Storage temperature:​ -20 to 70 °C
Functional Features​
Software-configurable switch topology​
High channel count in one module​
Reed relays​ for long life and fast operation
Multiplexer, matrix and independent bank modes​
Suitable for automated test routing​
Material And Structure​
Module:​ SCXI single-width switch module
Switch array:​ reed relay array on multilayer PCB
Front connectors:​ 176-pin
Backplane:​ SCXIbus communication
Working Principle​
Internal relays close or open signal paths​ under software control. The module routes selected input lines to measurement or source lines. The switching topology is configured in software rather than by hardware jumpers.
Application Scenarios​
Automatic test equipment​
Sensor matrix routing​
Multi-device scanning​
Production test systems​
Research lab signal routing​
High-channel switching cabinets​
Installation Requirements​
Install in compatible SCXI chassis
Use supported 176-pin cable and connector accessories​
Configure switch routes in NI Switch software or LabVIEW​
Ensure relay contact limits​ are not exceeded
Provide adequate ventilation
Usage Notes​
Do not​ switch loads above 60 VDC, 0.4 A or 10 W per channel​
Allow relay settle time​ for accurate scanning
Avoid hot-switching inductive loads​ without protection
Plan relay cycle life​ for long-term test systems
NI marks the model as discontinued