The Bender 107TL47 is an insulation monitoring device (ISOMETER®) that monitors the insulation resistance of single-phase and three-phase AC IT systems in medical locations. It also monitors the IT system transformer for overload and overtemperature. When combined with the MK2430 alarm indicator and test unit, it complies with DIN VDE 0100-710 and IEC 60364-7-710 requirements.
Description
Key Features:
-
Insulation Resistance Monitoring: Adjustable response value from 50 to 500 kΩ.
-
Load & Temperature Monitoring: Monitors IT system transformer for overload and overtemperature.
-
Connection Supervision: Continuously monitors measuring leads to system, earth, current transformer, and temperature sensors.
-
Alarm Indication: Alarm LED, common alarm relay with one potential-free changeover contact.
-
Backlit LC Display: Provides clear visual feedback and alarm messages.
-
Test Function: Integrated test button; external test button can be connected.
-
BMS-Bus Interface: For building management system integration.
-
Two-Wire Remote Communication: Transfers alarm messages to remote MK2430 units installed in medical locations.
Technical Specifications:
|
Parameter
|
Specification
|
|---|---|
|
Brand
|
BENDER
|
|
Model
|
107TL47
|
|
Product Type
|
Insulation Monitoring Device (ISOMETER®)
|
|
Monitored Systems
|
Single-phase AC and 3(N)AC IT systems
|
|
Adjustable Response Value
|
50…500 kΩ
|
|
Display
|
Backlit LC display
|
|
Interface
|
BMS-bus
|
|
Alarm Output
|
Common alarm relay, 1 potential-free change-over contact
|
|
Transformer Monitoring
|
Load and temperature monitoring
|
|
Testing
|
Test button (integrated); external test button connectable
|
|
Weight
|
0.58 kg
|
|
Origin
|
USA (Bender Inc.)
|
|
Compliance
|
DIN VDE 0100-710, IEC 60364-7-710
|
|
Warranty
|
12 months
|
Working Principle:
The device continuously measures insulation resistance between the monitored IT system and earth. When measured values exceed the limiting value (insulation resistance, load current, or temperature), an alarm is initiated—the alarm relay switches, the ALARM LED lights up, and a message appears on the LC display. The alarm message is transferred via a two-wire interface to remote MK2430 alarm indicators, immediately informing technical or medical staff.
Application Scenarios:
-
Medical IT Systems: AC and 3(N)AC IT systems in medical locations (operating rooms, intensive care units).
-
Transformer Monitoring: Overload and overtemperature monitoring of IT system transformers.
-
Critical Care Areas: Where continuous power availability and insulation integrity are paramount.
-
Three-Phase Systems: Requires CMS460-D4-2 load current monitor and STW2 current transformers for phase conductor current measurement.
Installation & Usage Notes:
-
For three-phase load current detection, pair with CMS460-D4-2 load current monitor and STW2 current transformers.
-
The highest value of the load current is evaluated by the electronics and made available at the input k/l of the 107TL47.
-
Measuring leads to system, PE, current transformer, and temperature sensor are continuously monitored; interruption or short-circuit triggers a message.
-
Function can be verified by pressing the integrated Test button.
-
Cable length for remote communication must not exceed 50 m.
-
Combine with MK2430 alarm indicator and test combination for full DIN VDE 0100-710 compliance.