Product Short Description

Product Introduction

LC100SSP7 is a high-precision four-electrode sanitary conductivity sensor designed for pharmaceutical, food and beverage pure water and clean-in-place (CIP) process monitoring. It eliminates polarization measurement errors via four-electrode technology and complies with global hygienic process industry standards for sanitary pipeline installation.

Description

Technical Specifications

  • Measurement Principle: Four-electrode potentiometric conductivity measurement
  • Conductivity Measuring Range: 0.01 μS/cm ~ 2000 mS/cm
  • Full-Scale Measurement Accuracy: ±0.5% of reading
  • Process Temperature Range: -20°C ~ +130°C (continuous); 143°C SIP steam sterilization peak
  • Maximum Process Pressure: 10 bar at 130°C continuous operation
  • Fitting Connection Type: SP7 1.5-inch sanitary tri-clamp ferrule connection
  • Ingress Protection Rating: IP67 sensor head housing
  • Electrode Material: 316L medical grade stainless steel
  • Insulator Body Material: Medical grade PEEK polymer
  • Cable Standard Length: 5 meters shielded signal cable
  • Compliance Certifications: EHEDG Type EL-A Class I, USP <645>, 3A sanitary standard

Function Features

  1. Four-electrode design eliminates cable resistance and polarization errors for ultra-pure water accurate measurement
  2. Crevice-free flush mounted sanitary body with no dead zones to prevent biofilm bacterial growth
  3. Compatible with CIP chemical cleaning and SIP saturated steam in-place sterilization cycles
  4. Integrated built-in temperature compensation sensor for automatic conductivity temperature correction
  5. Direct compatible with ABB Liquiline CM44x, CM42 transmitters for HART and Modbus digital signal output

Performance Parameters

  • Temperature compensation range: 0°C ~ 130°C automatic linear correction
  • Response time to conductivity step change: ≤2 seconds
  • Minimum detectable conductivity resolution: 0.001 μS/cm
  • Maximum allowable flow velocity across sensor head: 3 m/s
  • Electrode service life under standard hygienic operation: ≥5 years
  • Calibration stability interval: 12 months between factory calibration verification

Material Composition

  • Sensor electrode probes: 316L medical stainless steel
  • Main sensor body insulator: Unfilled food/pharma grade PEEK
  • Tri-clamp ferrule fitting: 316L stainless steel electro-polished surface
  • Internal sealing gaskets: EPDM sanitary rubber (FDA food contact compliant)
  • Outer cable sheath: PUR oil-resistant shielded cable
  • Cable shielding: Tinned copper braided full coverage shield

Structural Characteristics

  • Compact cylindrical flush-mounted sanitary sensor head without protruding electrode cavities
  • Integrated internal PT1000 temperature compensation element encapsulated within PEEK body
  • Single-piece electro-polished stainless steel tri-clamp SP7 connection base
  • Molded cable strain relief at sensor head to prevent wire fatigue breakage
  • Fully sealed internal electronics cavity isolated from process fluid

Working Principle

Two excitation electrodes inject low-amplitude alternating current into process fluid; two separate potential measurement electrodes capture fluid voltage drop independent of electrode polarization effects. Built-in PT1000 temperature sensor transmits fluid temperature data to transmitter to automatically normalize conductivity readings to standard 25°C reference temperature for regulatory compliance recordkeeping.

Advantage Highlights

  1. EHEDG and 3A certified hygienic design eliminating bacterial contamination risk in sterile production lines
  2. Four-electrode technology maintains high accuracy across ultra-pure water to concentrated brine measuring ranges
  3. Withstands repeated high-temperature SIP steam sterilization without permanent measurement drift
  4. IP67 sealed sensor head tolerates production line washdown environments
  5. Native digital communication compatibility with ABB Liquiline process transmitters for Industry 4.0 data logging

Applicable Industries

Pharmaceutical sterile water systems, food & beverage beverage blending, biotech bioreactor media monitoring, semiconductor ultra-pure water production, cosmetic sterile formulation lines

Installation Requirements

  1. Mount via standard 1.5-inch SP7 sanitary tri-clamp with FDA compliant EPDM gasket between sensor and pipeline ferrule
  2. Install sensor pipeline branch with minimum 5 pipe diameters upstream and downstream straight pipe to avoid flow turbulence measurement error
  3. Route signal cable with minimum 30cm separation from high-current motor power cables
  4. Single-point shield grounding at transmitter terminal cabinet only (no dual-end cable grounding)
  5. Mount sensor with cable exit pointing downward to prevent process condensate ingress into cable gland

Usage Precautions

  1. Avoid mechanical impact to PEEK sensor body; surface scratches create bacterial adhesion dead zones
  2. Do not exceed 143°C peak SIP temperature or 10 bar continuous process pressure ratings
  3. Calibrate sensor annually with certified traceable conductivity standard solutions
  4. Replace EPDM sealing gaskets every 12 months to prevent fluid leakage and microbial ingress
  5. Flush sensor with deionized water post CIP/SIP cycles to remove residual chemical cleaning agents

Reviews

There are no reviews yet.

Be the first to review “ABB LC100SSP7”

Your email address will not be published. Required fields are marked *

View Related Products